b
d
c
Name and detail specifications of this product
might differ in Europe, Middle East and Africa.
46
46
1
Notes for observation
· The objectives should be between 5X and 150X.
· The thickness of the observable sample should be between 0 and 9 mm
(MX-TILLB) or between 0 and 3 mm (MX-TILLA) including the thickness
6 mm of the glass plate (BH3-SPG6, MX-SPG1412).
· If the glass thickness exceeds the above mentioned range, the image
becomes dark due to insufficient NA of the illumination. If the objective
below 5X is used, insufficiency in the illumination field or in the peripheral
light intensity may result.
· If using the objective with higher NA than NA of the condenser (MX-TILLB:
0.6 or MX-TILLA: 0.5), the image becomes dark due to insufficient NA.
2
Adjusting the transmitted light illumination
1
Set the main switch a of the microscope frame to (ON), and rotate
the light intensity knob b counterclockwise completely.
a
2
Set the main switch c of the LED light source for light guide (LG-
LSLED) to (ON), and adjust the brightness using the light intensity
volume knob d .
TIP
Rotate the light intensity adjustment knob d clockwise
(toward the scale 5) to increase the brightness or rotate it
counterclockwise (toward the scale 0) to decrease the brightness.
For details, refer to the instruction manual provided with the LED
light source for light guide (LG-LSLED).