Problem
i) The system cannot be
controlled by PC.
Cause
The interface cable is not
connected.
The main switch is set to
while the communication between
PC and the microscope is in
process.
The device driver is not installed.
The device is not detected properly. Scan the device again with the
The USB cable is connected to the
USB connector of PC for the first
time.
The communication between
the microscope frame and PC is
disconnected.
Remedy
Connect the interface cable.
Close the application software.
(OFF)
Then, set the main switch to (ON)
and start the application software.
Install the device driver from
the installation media of the
application software.
device manager of the operation
system.
Install the device driver in
accordance with the installation
wizard.
Remove the interface cable
from PC and connect it again.
Then,restart the application
software.
If the communication is still not
restored,turn OFF the power
supplies of the microscope and PC
and turn ON again.
MX63/MX63L
Page
104
24, 104
24, 104
79
79