Setting the OP light path

If you need to attach the reflected light filter, attach it together with this setting. (See next page.)
Loosen the clamping screws on the top cover a using the Allen screwdriver (2 mm) to remove the top cover.
1
Rotate the observation light path selection knob b counterclockwise to remove it. (This selection knob should be
2
retained by the user because it will be reused when restoring the light path switching stroke to the original setting.)
Remove the hexagon socket head cap screws (4 x 16 ) d (4 positions) of the observation light path unit c using
3
the Allen wrench (3 mm), and place the unit c upside down.
Loosen the mirror unit clamping screw e using the Allen screwdriver, insert the desired mirror unit f into the mount
4
dovetail (so that the product name on the side of the unit is upside down), and tighten the clamping screw e .
Remove the screw g that limits the light path selection stroke using the Allen wrench (3 mm). (This screw should be
5
retained by the user because it will be reused when restoring the light path switching stroke to the original setting.)
Return the observation light path unit c to the original position and secure it, and also secure the top cover a .
6
NOTE
Screw the provided longer light path selection knob firmly into the b position.
7
110
110
f
e
Be sure to use the Allen wrench (3 mm) to secure the observation light path unit c . Do not use the
Allen screwdriver.
c
g
d
b
a
d