Olympus MX63 Instructions Manual page 40

Semiconductor/fpd/industrial inspection microscopes
Table of Contents

Advertisement

34
34
a
a
b
Procedure to replay the brightness
1
Press the LIM switch a to set the "Replay" mode. (LIM switch a
is ON.)
2
When you select the desired objective or the observation method, the
stored brightness is set automatically.
Restoring the stored brightness (LIM function)
to the factory default setting
1
Set the main switch of the microscope frame to
2
Set the main switch of the microscope frame to (ON) while pressing
both the LIM switch a and the SET switch b . Keep pressing the
LIM switch a and the SET switch b . The buzzer is heard three
times after 5 seconds, and the stored brightness returns to the factory
default setting.
· If you release the LIM switch and the SET switch within 5
TIP
seconds, the stored brightness (LIM function) does not return to
the factory default setting and the microscope starts normally.
· The buzzer sound can be set to silent by setting the DIP
switches. For setting procedures, see "Setting the DIP switch /
AS.PRESET switch" (P.98).
3
Set the main switch of the microscope frame to
(OFF).
(OFF).