7. Error indications by lighting and blinking of buttons
Front panel
"1" and "2" blink. g) Both two objective
selection buttons keep
blinking.
"1" to "4" are OFF. h) The buttons from "1" to
"4" keep blinking.
Repair request
If you cannot solve the problems even though taking actions described in Troubleshooting, please contact Olympus
for assistance. At that time, please tell them the following information as well.
· Product name and abbreviation (Example: Stage MX-SIC8R)
· Product number
· Problem
Problem
1
2
3
The DIP switch is not set
properly.
The objective interferes with
the sample or the stage.
The connection between
the revolving nosepiece
and the microscope frame
is disconnected.
The revolving nosepiece is
damaged.
The hand switch (BX3M-HS)
or the hand switch for
exposure(U-HSEXP) is
unpluged after turning
ON the main switch of
microscope.
Cause
4
Set the DIP switch properly.
Make sure that the objective
does not interfere the sample or
stage, and turn OFF the power
supplies of the microscope and
turn ON again.
Attach the revolving nosepiece
again properly.
Contact Olympus.
Set the main switch to
and connect the cords securely.
MX63/MX63L
Remedy
Page
98
65
(OFF)
96
81
81