Setting the analyzer and the polarizer for
1
transmitted light illumination
When (OP) is selected for observation light path
1
When DF is selected for observation light path
1
a
2
c
3
4
5
6
7
Insert the polarizer (U-PO3) a in the polarizer insertion slot.
Polarizer insertion position
First level (pulled out)
Second level (pressed in)
If the DIC slider is engaged in the light path, remove it from the light path.
Engage the 10X objective or 20X objective in the light path and bring
the sample into focus approximately.
If the cover or the dummy slider is attached to the analyzer insertion
slot or the polarizer insertion slot, remove it.
Insert the analyzer (U-AN360-3) a in the analyzer insertion slot.
Analyzer insertion position
First level (pulled out)
Second level (pressed in)
Rotate the analyzer rotation dial to set to 0°.
Insert the polarizer (U-PO3) b in the polarizer insertion slot.
Analyzer insertion position
First level (pulled out)
Second level (pressed in)
Rotate the analyzer rotation dial c so that the observation field of
view becomes the darkest (crossed nicols).
Light path
Empty hole
Polarizer
Light path
Empty hole
Analyzer
Light path
Empty hole
Polarizer