Using The Filters For Transmitted Light Illumination - Olympus MX63 Instructions Manual

Semiconductor/fpd/industrial inspection microscopes
Table of Contents

Advertisement

b
d
a
Name and detail specifications of this product
might differ in Europe, Middle East and Africa.
f
g
h
5

Using the filters for transmitted light illumination

Filters can be inserted to three positions described below.
Note, as the filter at the tip of the light guide is a built-in type, it is
recommended to attach the constantly-used color temperature conversion
e
filter 25LBD (for converting the illumination light to the daylight color).
Filter insertion slot a (one filter):
A filter same as the filter for reflected light illumination and the polarizer
(U-PO3) can be inserted.
Tip of light guide b (one filter):
25LBD-IF, 25ND6 or 25ND25 is attachable.
Inside the light source c (one filter):
LBA filter is attachable. For details, refer to the instruction manual
c
provided with the light source (LG-LSLED).
Attaching the filter at the tip of light guide
Loosen the clamping screw d of the light guide holder using the Allen
1
screwdriver, and remove the light guide holder by holding the knob e .
Hold the filter holder f at the tip of the light guide with fingers and
2
rotate the filter holder counterclockwise to remove it. Attach the required
filter g and rotate the filter holder clockwise to secure it.
If you need to remove the light guide, loosen the clamping screw h .
3
MX63/MX63L
49
49

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mx63l

Table of Contents