Attaching the focus aid (MX-FA)

a
Loosen the clamping screws on the top cover a using the Allen wrench (2 mm) to remove the top cover.
1
Peel off the sticker b of the focus aid knob insertion slot.
2
Push the focus aid unit in back-left direction (arrow direction d in the picture) facing from the right side of the
3
microscope, and secure it using the clamping screws e (2 positions).
Clean the place to attach the sticker f for focus aid on the side panel using the cleaning paper, etc. moistened with
4
absolute alcohol, and attach the sticker using the mount hole as a guide.
Insert the focus aid knob g from the insertion slot and rotate clockwise to attach it.
5
6
Attach the top cover.
112
112
b
e
d
g
f