Attaching the revolving nosepiece
Only U-D5BDREMC, U-D6REMC, U-P5REMC,
NOTE
U-P5BDREMC and U-D6BDREMC are available with this
microscope.
1
Rotate the coarse focusing knob to lower the stage sufficiently.
Loosen the revolving nosepiece clamping screw a using the Allen
2
screwdriver provided with microscope frame.
If the clamping screw is loosened too much, the revolving
NOTE
nosepiece is caught with the clamping screw and
it cannot be attached. And, if the clamping screw is
loosened furthermore, the clamping screw may come off.
1
2
If
and
in "Removing the revolving nosepiece" (P.65) are
1
performed,
and
3
Insert the revolving nosepiece from the front side by aligning the
slide dovetail b of the revolving nosepiece with the nosepiece
mount dovetail c of the microscope frame, and push it until it
touches the end.
· When attaching the revolving nosepiece, lower the stage
NOTE
sufficiently so that objectives do not touch the stage.
· Push in the revolving nosepiece mount dovetail
completely until it touches the end. If the clamping screw
is tightened without inserting the revolving nosepiece
completely, the mount section may be deformed and the
revolving nosepiece cannot be attached or removed. Also,
the connection failure of the connector may occur.
· The revolving nosepiece is heavy. Attach it carefully not to
drop it.
4
While pushing the revolving nosepiece in the attaching direction with
your left hand, hold the Allen screwdriver between the pointing finger
and the thumb and tighten the revolving nosepiece clamping screw
a to secure the revolving nosepiece.
5
Set the DIP switches of the attached revolving nosepiece when all
assemblies are completed. For setting procedures, see "Setting the
DIP switch / AS.PRESET switch" (P.98).
MX63/MX63L
2
are not necessary.
67
67