Attaching the reflected light filter (26ND0.5)

g
e
f
Loosen the clamping screws on the top cover a using the Allen screwdriver (2 mm) to remove the top cover.
1
Rotate the observation light path selection knob b counterclockwise to remove it. (This selection knob should be
2
retained by the user because it will be reused when restoring the light path switching stroke to the original setting.)
Remove the hexagon socket head cap screws (4 x 16 ) d (4 positions) of the observation light path unit c using
3
the Allen wrench (3 mm), and place the unit c upside down.
Remove the screws f (2 positions) that clamp the filter frame e using the Phillips screwdriver and remove the filter
4
frame e .
Insert the filter g into the filter holding slit (long slit) of the filter frame.
5
The filter frame is slightly deformed toward outside, but this is normal.
TIP
Secure the filter frame e at the original position.
6
c
d
b
MX63/MX63L
a
d
111
111