Setting the DIP switch / AS.PRESET switch

Setting the DIP switch
Make sure that the connections are correct.
Before setting the DIP switch a , set the main switch of the microscope
to
(OFF). The switch settings are read and defined only when the power
is turned ON.
Be careful not to touch the internal circuit boards when
NOTE
a
setting the DIP switches. If you touch it, the internal circuit
boards may be damaged by static electricity.
Since the human body is charged with a small amount of
CAUTION
static electricity, discharge it from your body before setting
switches. You can discharge the static electricity by lightly
touching any metallic objects around you with hand.
DIP
a
switch
ON OFF
Revolving nosepiece type
(Number of holes)
Buzzer sound
Aperture diaphragm open/
close button setting
Glare prevention by AS
when changing objectives
Glare prevention by LED
when changing objectives
Reserved by maker
Function
U-D5BDREMC/U-P5REMC/
OFF
U-P5BDREMC
ON U-D6REMC/U-D6BDREMC
OFF The buzzer sound is heard.
ON The buzzer sound is not heard.
OFF The setting is enabled.
ON The setting is disabled.
OFF Does not prevent the glare.
ON Prevents the glare.
OFF Prevents the glare.
ON Does not prevent the glare.
OFF This switch is always kept OFF.
ON
: Factory default setting
Setting