CAUTION - Lamp (halogen bulb and mercury burner) -
4 4
Remove the power cord from the product when replacing the
lamp.
To avoid electric shock hazards and burns when replacing the lamp,
set the main switch to
the power supply in advance. When you replace the lamp immediately
after use of the microscope, wait until the lamp housing and the lamp
are sufficiently cooled down.
[For U-LH100L-3]
12V100WHAL-L (made by PHILIPS Co. 7724I)
12V100WHAL (made by PHILIPS Co. 7023)
[For U-LH100IR]
Applicable lamp
12V100WHAL-L (made by PHILIPS Co. 7724I)
[For U-LH100HGAPO, U-LH100HG]
USH-103OL (made by Ushio Co.)
[For U-HGLGPS]
SHI-130OL (made by Ushio Co.)
Secure the sufficient space when installing the lamp housing.
The surface of the lamp housing on the back of the microscope
gets very hot. Therefore, when installing the microscope, secure the
sufficient space (100mm or more) around the lamp housing, particularly
on the top and bottom surfaces. Also, if the power cord or cables
contact the lamp housing, they may melt and cause electric shock. Be
sure to lay out the power cord and cables sufficiently away from the
lamp housing.
· As the UV ray of the mercury burner is harmful to your eyes, do not
light the mercury burner when the lamp housing is not attached to the
microscope.
· The lifetime of the illumination device is influenced significantly by
operating conditions (ambient temperature/humidity, voltage of the
power supply, lighting time, etc.), but is approximately eight(8) years
or 20,000 illumination hours if operating it 8 hours a day, whichever is
shorter, as a rough standard.
· Since the lamp generates a high heat, perform the inspection
according to "12 Preventive inspection sheet for illumination devices"
on page 106. If you find the unexpected phenomena, e.g. smoke, etc.
during operation, turn OFF the power immediately and contact
Olympus.