Olympus MX63 Instructions Manual page 75

Semiconductor/fpd/industrial inspection microscopes
Table of Contents

Advertisement

d
e
g
d
a
2
4
While pressing down the bulb fixing lever d , hold the halogen bulb
e wrapped with gauze, etc. and insert the terminal f until it touches
the pin position g .
Place the bulb fixing lever carefully to the original position to secure
the bulb.
Be careful not to touch the bulb directly with your hand.
NOTE
f
If fingerprints or stains are attached to the bulb, wipe it
with a soft cloth, etc. to prevent the lifetime reduction or
the bulb explosion. If fingerprints or stains are attached to
the bulb, clean it by wiping gently with a piece of cleaning
paper slightly moistened with absolute alcohol.
5
Fit the halogen lamp housing from the above and while pressing down
the mounting screw, tighten it.
3
Attaching the halogen lamp housing
Loosen the mounting screw a (2 positions) on the microscope frame
1
using the Allen screwdriver.
2
Insert the halogen lamp housing in the light source mount hole until
it touches the end.
Attach the lamp housing with the heat radiating fin b facing
CAUTION
up, and secure he sufficient space above, below and at rear
of the lamp housing. Otherwise, fire may be caused.
3
Tighten the mounting screw using the Allen screwdriver.
Do not tighten the screws too firmly. The lamp housing
NOTE
mount may be deformed.
4
Connect the cable for halogen lamp housing to the connector of the
power supply (TH4-100/TH4-200). For details, see "Connecting the
cables" (P.96).
Applicable lamp
JCR12V100WB (made by Ushio Co.)
Replacing the light guide halogen bulb for
1
transmitted light illumination
For procedures to replace the light guide halogen bulb
for transmitted light illumination, refer to the instruction
manual provided with the unit.
MX63/MX63L
69