Safety Precautions - Olympus MX63 Instructions Manual

Semiconductor/fpd/industrial inspection microscopes
Table of Contents

Advertisement

Safety precautions

If the product is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the product
may also be damaged.
Read this instruction manual carefully and always use this product according to this instruction manual.
The following symbols are used in this instruction manual.
CAUTION
NOTE
CAUTION - Installation of the product -
Install the product on a sturdy, level table or bench.
· If you install the microscope on an unstable table or a tilted table, the microscope may topple and you may get
injured.
· Install the product on an desk where the tilt of the top surface of the desk meets following conditions.
Otherwise, the stage may move spontaneously.
・ MX63 : 1° or less
・ MX63L : 20' or less
· For safety, do not place a mat, etc. under the device.
Use of the adjuster.
The center of gravity changes depending on the sample weight, sample position or stage movement, etc.
In particular, as MX63L uses a large-size stage, the center of gravity changes significantly.
To prevent the microscope from unexpected overturning, adjust the adjuster at the bottom of the microscope to take
the overturning prevention actions.
For adjustment procedures, see "Use of the adjuster" (P.102).
Seismic protection
In order to prevent the microscope from overturning in case of seismic, etc., use the L-shape fixture to take the
reinforcement countermeasures.
For the seismic protection, see "Seismic protection" (P.102).
2 2
: Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury.
: Indicates a potentially hazardous situation which, if not avoided, may result in
damage to the equipment or other property.
: Indicates the useful knowledge or information for use.
TIP