Setting the DIC slider

1
Loosen the mounting knob a at the rear right of the revolving
nosepiece and pull out the dummy slider b .
2
Insert the DIC slider c for reflected light observation in the slider
insertion slot of the revolving nosepiece with the display surface facing
up, and stop inserting at the first level (when the clicking sound is
heard at first).
DIC slider position
First level (pulled out)
Second level (pressed in)
3
Tighten the mounting knob a to secure the slider.
Only with the DIC slider U-DICR , slide the selection lever d depending
4
on the objective in use.
Position of light path
selection lever d
Pressed in
Pulled out
With the U-DICRH or U-DICRHC without selection lever, the applicable
objectives are as follows.
DIC slider c
U-DICRHC
U-DICRHC
Light path
OUT
IN
Applicable objectives
MPLFLN/MPLFLN-BD series
UIS2
MPLAPON series
UMPlanFI/UMPlanFI-BD series
UIS
MPlanApo20X, 100X
MPlanApo100XBD
UIS2
LMPLFLN/LMPLFLN-BD series
LMPlanFI/LMPlanFI-BD series
UIS
LMPlanApo/LMPlanApo-BD series
Applicable objectives
MPLFLN/MPLFLN-BD series
UIS2
MPLAPON series
UMPlanFI/UMPlanFI-BD series
UIS
MPlanFI-BD series
MPlanApo20X, 100X
UIS2
LMPLFLN/LMPLFLN-BD series
LMPlanFI/LMPlanFI-BD series
UIS
LMPlanApo/LMPlanApo-BD series