Main Switch - Olympus MX63 Instructions Manual

Semiconductor/fpd/industrial inspection microscopes
Table of Contents

Advertisement

3-2 Main switch

a
24
24
Rotate the light intensity knob a counterclockwise fully and set the
1
main switch b to (ON).
In case of emergency, disconnect the power cord from
CAUTION
the product , and set the main switch to
down the power supply. Do not place any object near the
main switch so that the main switch operation will not be
interfered.
· When the power is turned ON, the buzzer sound is heard
TIP
once. The buzzer sound can be set to silent by setting the DIP
b
switches. For setting procedures, see "Setting the DIP switch /
AS.PRESET switch" (P.98).
· After the power is turned ON, the aperture diaphragm starts
initializing and the aperture diaphragm open/close button on
the front panel blinks for several seconds.
2
If following units are combined, set the main switch of the respective
unit to (ON).
Combinable units
Mercury lamp housing
Halogen lamp housing
Light source
LED light source for light
guide
* In general, the arc image becomes stable 5 to 10 minutes after
turning ON the power supply, but the discharge type mercury burner
may not turn ON by one operation due to its nature. In this case,
set the main switch to
it to  (ON) again.
** Name and detail specifications of this product might differ in Europe,
Middle East and Africa.
Unit to set to ON
Main switch of the power supply for
mercury burner (U-RFL-T)*
Main switch of the power supply for
halogen bulb (TH4-100/TH4-200)
Main switch of the light source
(U-HGLGPS)
Main switch of the LED light source
for light guide (LG-LSLED)**
(OFF) once, wait 5 to 10 seconds and set
For details, refer to the instruction manual provided
with the respective unit.
(OFF) to shut