Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User's Manual
7.3 A/D REFERENCE Calibration
■
To calibrate the HV OUTPUT and RTN/LOW terminals of a channel, conduct the
following calibration procedure without any UUT or cables connected.
A/D REFERENCE Calibration Screen:
CALIBRATION
A/D REFERENCE
CHANNEL: 1
VOLTAGE
1.500V
1.
Press START to do the A/D REFERENCE calibration.
2.
The A/D REFERENCE calibration is done when the status indicator shows PASS.
3.
Press ▼ to go to the ACV 5kV AGC OFFSET (0.1kV) calibration.
7.4 Voltage Calibration
7.4.1 ACV Calibration
■
Connect an ACV high voltage meter to the tester with the high voltage terminal
connected to the HV OUTPUT and the low voltage terminal connected to the
RTN/LOW of the channel to be calibrated.
ACV 5kV OFFSET (0.1kV) Calibration Screen:
CALIBRATION
ACV 5kV AGC OFFSET (0.1kV)
CHANNEL: 1
OUTPUT
0.100kV
STANDARD
0.102kV
DEF=0.1kV
7-4
CURRENT
1.500V
REMOTE LOCK
READING
0.101kV
REMOTE LOCK
NEXT CH
PREV CH
CORR
ERROR
NEXT CH
PREV CH
CORR
ERROR