A/D Reference Calibration; Voltage Calibration; Acv Calibration - Chroma 19020 User Manual

Multi-channel sync hipot tester
Hide thumbs Also See for 19020:
Table of Contents

Advertisement

Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User's Manual

7.3 A/D REFERENCE Calibration

To calibrate the HV OUTPUT and RTN/LOW terminals of a channel, conduct the
following calibration procedure without any UUT or cables connected.
A/D REFERENCE Calibration Screen:
CALIBRATION
A/D REFERENCE
CHANNEL: 1
VOLTAGE
1.500V
1.
Press START to do the A/D REFERENCE calibration.
2.
The A/D REFERENCE calibration is done when the status indicator shows PASS.
3.
Press ▼ to go to the ACV 5kV AGC OFFSET (0.1kV) calibration.

7.4 Voltage Calibration

7.4.1 ACV Calibration

Connect an ACV high voltage meter to the tester with the high voltage terminal
connected to the HV OUTPUT and the low voltage terminal connected to the
RTN/LOW of the channel to be calibrated.
ACV 5kV OFFSET (0.1kV) Calibration Screen:
CALIBRATION
ACV 5kV AGC OFFSET (0.1kV)
CHANNEL: 1
OUTPUT
0.100kV
STANDARD
0.102kV
DEF=0.1kV
7-4
CURRENT
1.500V
REMOTE LOCK
READING
0.101kV
REMOTE LOCK
NEXT CH
PREV CH
CORR
ERROR
NEXT CH
PREV CH
CORR
ERROR