II 3420

MAMMOGRAPHY CASSETTE PREPARATION.

All MINILOADER CASSETTES must be patched correctly otherwise reliable operation is impossible. The
CASSETTES should be patched as shown below using TOOLS 9194531 & 9194511. The correct PATCHES
are PART NUMBER 9194551, REFLECTIVE FOIL 8 x 15 mm. Place 3 PATCHES (8 x 15 mm) as shown
in figures 23 and 24, using the TOOLS as indicated.
If DIPSWITCH 6 on PCB 401 is turned on, no MAMMOGRAPHY CASSETTES should have a second
(TYPE 2) patch.
This is the TYPE 2
position.
Figure 23
Figure 24
© Kodak Ltd.
PATCH 2
PATCH 3
Page II-20
INSTALLATION INSTRUCTIONS
PATCH 1
January 1995