GE LOGIQ E9 Service Manual page 95

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
3-6-9
Complete power down (cont'd)
3-6-9-1
Sleep Mode
Only available in LOGIQ E9 running R2.x.x and R3.x.x software.
Use Sleep Mode when you do a portable exam in order to reduce the time to start up the system. When
you use Sleep Mode, it only takes 90 seconds to start up the system versus 2-3 minutes. To activate
Sleep Mode, press the ON/OFF button and select Sleep.
For optimum system operation, it is recommended that a full shutdown of the system is performed at
NOTICE
least once every 24-hour period. If you shut down the system at the end of the day, no other action is
needed.
Before returning a system to the customer, perform the functional checks. See:
checks" on page
. 2
EV
4-31.
Chapter 3 LOGIQ E9 Setup
LOGIQ E9 S
M
ERVICE
ANUAL
Section 4-3 "Functional
3 - 31

Advertisement

Table of Contents
loading