GE LOGIQ E9 Service Manual page 451

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
8-6-3-2
LCD Monitor V2 Arm assembly replacement (used in R4.x production)
Table 8-67
1.
Remove the LCD Monitor.
2.
WHEN REMOVING THE LCD ARM, KEEP
THE ARM LOCKED, AS SHOWN TO
ENSURE THE LCD ARM ASSEMBLY IS
IN THE LOCKED POSITION UNTILL THE
ARM AND LCD ARE INSTALLED. THE
SPRINGS TO SUPPORT THE LCD CAN
CAUSE THE ARM TO SPRING OPEN
CAUSING SEVERE PERSONAL INJURY
AND PROPERTY DAMAGE.
3.
Remove the Bulkhead Cover.
4.
Disconnect the LCD cables; video (P10)
and power (P3) from the LCD.
The Video Cable has a retainer and screw.
Loosen the screw with a small, flat blade
screwdriver.
. 2
EV
LCD Monitor V2 Arm assembly removal - R4
Steps
WARNING
WARNING
Chapter 8 Replacement procedures
LOGIQ E9 S
ERVICE
Corresponding Graphic
M
ANUAL
8 - 105

Advertisement

Table of Contents
loading

Table of Contents