GE LOGIQ E9 Service Manual page 790

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
10-6-3
Leakage Current Limits (cont'd)
NOTE:
Leakage current limits for Ultrasound system operation on 230-240 Volt mains
Table 10-11 ISO (on Dale 600) and Mains Applied (on Dale 601) Limits*
*ISO (on Dale 600) and Mains Applied (on Dale 601) refer to the sink leakage test where mains (supply)
voltage is applied to the part to determine the amount of current that will pass (or sink) to ground if a
patient contacted mains voltage.
10 - 12
. 2
EV
Probe Type
BF
CF
Section 10-6 - Electrical Safety Tests
Measurement
5.0 mA (5000 µA)
0.05 mA (50 µA)
LOGIQ E9 S
M
ERVICE
ANUAL