GE LOGIQ E9 Service Manual page 428

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
8-5-20-4
Covers under XY / Frogleg motors removal (cont'd)
4.) Pull down and slide the cover(s) away from the XY / Frogleg mechanism. Be sure to flex the plastic
slightly so the plastic clears the XY.
5.) Disconnect the cable.
6.) Remove the screw securing the ground.
7.) Thread the cover(s) off the cable.
8-5-20-5
Covers under XY / Frogleg motors installation
1.) Thread the cover(s) on the cable.
2.) Perform a dry fit of the covers (confirm the covers face the correct way) before connecting the cable
and ground.
3.) Install the screw securing the ground.
4.) Connect the cable.
5.) Slide cover(s) into place.
6.) Replace the screws to the cover(s).
8 - 82
. 2
EV
Figure 8-37 Pull down and slide XY / Frogleg cover out