Capturing Network Logs with Network Sniffer (Software R2.x.x or later) (cont'd)
Use filters to limit the captured data. Filter by IP or by IP and port number.
By IP only: press on Capture Filter.
4.) Select IP only.
Use the IP address of the Scanner or the one from the DICOM device under test, on the Filter string
field, using the following syntax: (example) host 3.62.12.33 as shown in
Only" on page
. 2
EV
7-13, press OK.
Figure 7-14 Capture Filter IP Only
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 S
M
ERVICE
Figure 7-14 "Capture Filter IP
ANUAL
7 - 13