GE LOGIQ E9 Service Manual page 85

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
3-6-6-2
Connect a probe
NOTE:
It is not necessary to turn OFF power to connect or disconnect a probe. However, it is a good
idea select a different probe or to freeze the image when removing a probe to avoid
disconnecting a live probe.
DO NOT ALLOW THE PROBE HEAD TO HANG FREELY. EXCESSIVE IMPACT TO THE PROBE
CAUTION
WILL RESULT IN IRREPARABLE DAMAGE.
TO PREVENT PROBE CONNECTOR PINS DAMAGE, OR PCB BOARD DAMAGE, DO NOT USE
CAUTION
EXCESSIVE FORCE WHEN CONNECTING THE PROBES.
1.) Before connecting the probe:
a.) Perform a visual check of the probe pins and LOGIQ E9 sockets.
b.) Remove any dust or foam rests from the probe pins.
c.) Verify the probe and the probe cable for any visual damage.
2.) Hold the probe connector vertically with the cable pointing upward.
3.) Turn the connector locking handle counter-clockwise to the horizontal position.
4.) Align the connector with the probe port and carefully push into place.
5.) Turn the locking handle clockwise to the full vertical position to lock in place.
6.) Position the probe cable so that it is not resting on the floor.
KEEP THE PROBE CABLES AWAY FROM THE WHEELS.
CAUTION
DO NOT BEND THE PROBE CABLES.
DO NOT CROSS CABLES BETWEEN PROBES.
3-6-6-3
Disconnect a probe
1.) Select a different probe or Freeze the image before removing a probe in order to avoid
disconnecting a live probe.
2.) Rotate the lock handle counter-clockwise to the horizontal position to unlock the connector.
3.) Remove the connector from the port.
4.) Ensure that the probe head is clean before placing the probe in its storage case, see:
"Cleaning" on page 10-9
REFER TO THE TEE PROBE MANUAL FOR FURTHER INSTRUCTIONS (DIRECTION KZ192871).
CAUTION
. 2
EV
for cleaning instructions.
Chapter 3 LOGIQ E9 Setup
LOGIQ E9 S
M
ERVICE
ANUAL
10-4-10
3 - 21