IRECTION
7-5-17-15
Distinct Network Monitor R1.x.x
The Distinct Network Monitor has a sniffer that monitors network traffic and allows you to capture
network data without redirecting or altering it.
For directions on how to configure the sniffer go to
(Software R1.x.x)" on page
LOGIQ E9 R2.x.x or later Distinct was replaced by Wireshark. This menu is no longer available. Use
Alt+N to open the sniffer application.
7-5-17-16
Virtual Console Observation (VCO)
VCO is used by a remote service technician or the Online Center (OLC) to access and modify all
scanner settings and programs on the customer's ultrasound scanner.
7-5-17-17
Telnet Server Control
Telnet Server must be running to allow remote access to the DOS command prompt window. Click the
Start button to run Telnet Server.
NOTE:
Telnet Server is a Windows service that runs in the background. StopTelner Server when you
are not using it to perform a service action. It can slow the system down if left running in the
background.
7-5-18
Utilities - Scanner Utilities
7-5-18-1
DICOM Verify
This utility provides an easy way to verify DICOM connectivity between the scanner and DICOM devices
on the network.
1.) Enter AE Title, IP Address, and Port values of the DICOM device.
2.) Check the Loop checkbox to repeat the operation, or leave it unchecked to perform the operation
once.
3.) Click the Verify button to see the results.
4.) Uncheck the Loop checkbox to stop the operation.
7-5-18-2
Replacement
Field is not populated on the LOGIQ E9.
7-5-18-3
PM
Field is not populated.