Mechanical safety
While the software install procedure is designed to preserve data, you should save any
WARNING
WARNING
patient data, images, system setups to a DVD or hardcopy before doing a software
upgrade.
PRIOR TO ELEVATING THE SCANNER, VERIFY THAT THE KEYBOARD IS LOCKED IN
WARNING
WARNING
ITS LOWEST POSITION. VERIFY THAT THE FRONT BRAKE IS LOCKED AND THE
SCANNER IS UNABLE TO SWIVEL. VERIFY THAT THE REAR BRAKES ARE IN THE
LOCKED POSITION.
WHEN THE UNIT IS RAISED FOR A REPAIR OR MOVED ALONG ANY INCLINE, USE
WARNING
WARNING
EXTREME CAUTION SINCE IT MAY BECOME UNSTABLE AND TIP OVER.
ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN
WARNING
WARNING
EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND
PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR
DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN
SERIOUS INJURY AND EQUIPMENT DAMAGE.
REMEMBER: If the front caster swivel lock is engaged for transportation, pressing the
WARNING
WARNING
release pedal once disengages the swivel lock. You must depress the release pedal a
second time to engage the brake.
The system should NOT be moved with the Operator I/O Panel extended. Move the
WARNING
WARNING
Operator I/O Panel to its centered and locked position. Lower the Operator I/O Panel as
much as possible before moving the system.
BEFORE YOU MOVE OR TRANSPORT THE SYSTEM, MAKE SURE TO LOCK THE LCD
CAUTION
MONITOR ARM FIRMLY AND FLIP DOWN THE MONITOR TO PREVENT DAMAGE TO THE
SYSTEM.
Always lock the Top/Upper Console in its parking (locked) position before moving the scanner
CAUTION
around.
TO AVOID INJURY WHEN YOU MOVE THE LCD MONITOR AND THE MONITOR ARM, DO NOT
CAUTION
PUT YOUR FINGER, HAND, OR OBJECT ON THE JOINT OF THE MONITOR OR THE MONITOR
ARM.
. 2
EV
Chapter 1 Introduction
LOGIQ E9 S
M
ERVICE
ANUAL
1 - 17