Top Console description
The Top Console includes:
LCD monitor
Operator Panel with;
-
-
-
-
-
speakers for stereo sound output (used during Doppler scanning/replay)
A flexible harness of electrical wires secures the connection between the Top Console and the rest of
the LOGIQ E9.
The Top Console can be moved up/down (Z-axis), sideways to the left and to the right (X-axis) and back
and forth (Y-axis). The XYZ Mechanism is commonly referred to as the frogleg.
The up/down movement of the console is motor driven. The control buttons for the electrical motor are
centered between the handles. A gas spring inside the LOGIQ E9 counterbalances the Z-axis (vertical)
movement, lessening the load the Z motor must move.
The brakes used for locking the console's X and Y movement are all operated by electrical motors
5-3-3-1
Top Console's location in the LOGIQ E9
The Top Console is located on the top of the LOGIQ E9, and includes the Main LCD monitor, the Monitor
Arm, the Operator Panel with the Touch Screen, Speakers, an alphanumeric keyboard, and frogleg
controls.
Figure 5-7 Top Consoles location (A) - R1 through R3 (left) and R4 shown
. 2
EV
An On/Off switch
A Touch Panel Screen and a Control Panel with controls for manipulating the picture quality
and for use in Measure & Analyze (M&A)
An alphanumeric keyboard (QWERTY keyboard).
Trackball
XYZ Control, e.g., frogleg controls and lock to move the Top Console left/right, forward/
backward, and up/down.