GE LOGIQ E9 Service Manual page 140

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
Table 3-30
4.
Capture Restart clears the current traffic display, in preparation for the next DICOM exchange.
In the Wireshark Info column, entries with "Reject" or "Abort" definitely indicate issues. For example,
the above exchange occurred when the tried to send an exam to Viewpoint, but Viewpoint did not
have the correct AE Title for the .
3 - 76
. 2
EV
Network "Sniffer"
Network Connection: Capturing Wireshark
In the "Filter" bar, enter "dcm" or "dicom." (Which one Wireshark accepts, depends on its
version. Wireshark will indicate whether it accepts one or the other by turning the Filter
background green.) Press Apply.
Once Wireshark displays DICOM traffic, select an individual line and expand the DICOM tree
in the lower frame to view DICOM-specific data.
Section 3-9 - Connectivity Setup and Tips
Wireshark Info column
(continued on next page)
LOGIQ E9 S
M
ERVICE
ANUAL

Advertisement

Table of Contents
loading

Table of Contents