3-9-4-3
Sending Images and Data to a DICOM Image Storage Service
The provides three ways to send images and data:
1.) Dataflow
2.) Print Button
3.) Send To
1.) Dataflow
The default setup stores images in its local Hard Disk Drive. A dataflow allows the to send images to
other locations as well.
To setup a dataflow, follow these steps:
Table 3-12
1.
2.
3.
4.
. 2
EV
Dataflow
Utility
Navigate to Utility -> Connectivity -> Dataflow; press Add. The system creates a dataflow
called "NewDataflow."
Change the name to one of your choosing.
Notice the list of available services to the left of the << and >> buttons. Select each DICOM
service you want to attach to your dataflow and press >>. This makes it appear in the list on
the right, indicating that the service is now assigned to the dataflow.
NOTE: You cannot remove "Local Archive – Int HD" from the dataflow. This guarantees that
images will be saved on the local Hard Disk Drive of the and not just transmitted to the service.
Check Direct Store if you want an image generated by a print button to immediately be sent
to the dataflow's attached services rather than waiting until the exam is ended for all images.
NOTE: If you plan to generate long cine loops, "Direct Store" is NOT recommended (either on
the dataflow or via a Print button directly), because system response may be slow.
Chapter 3 LOGIQ E9 Setup
->
->
Connectivity
Dataflow screen
(continued on next page)
LOGIQ E9 S
M
ERVICE
ANUAL
3 - 49