GE LOGIQ E9 Service Manual page 257

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
7-3-5
Capturing Network Logs with Network Sniffer (Software R1.x.x) (cont'd)
10.)In the Filter to Apply field, select "My Computer".
11.)Select Close. The Configuration screen closes.
12.)On the Distinct Network Monitor Screen, select Capture > Packets and Statistics (Play).
NOTE:
The Network Sniffer will now collect data until you select Capture > Stop.
13.)Double-click on the Go Ichiro icon to start the system application.
14.)Perform the DICOM transaction you want to troubleshoot (such as Worklist Query, Send to PACS,
etc.). Perform the function several times to ensure the data is captured in the sniff.
. 2
EV
Figure 7-8 Capture Tab on Configuration Screen
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 S
M
ERVICE
ANUAL
7 - 9