Table 8-123 BEP installation - R3.x and earlier
4.
Reach inside the BEP and connect:
• the BEP cable to the Backplane J3 on a
GFI Configuration.
• the BEP cable to Backplane J2 (left) on a
MRX Configuration.
5.
NOTE: If you DO NOT reconnect the internal BEP cable to the Backplane, the system will power
up but WILL NOT scan or WILL LAUNCH into simulator mode.
6.
Insert the bottom lip of the BEP cover inside the base of the BEP.
7.
Connect the Battery cable to J3 on the EPS and EPS to BEP cables to the EPS.
8.
MAKE SURE all cables are clear of the BEP Cover and tilt the top of the BEP cover toward the
BEP.
Be careful not to pinch any of the cables when installing the BEP cover.
9.
Tighten the thumb screw or screws at the top of the BEP cover that attach the cover to the BEP.
. 2
EV
Steps
NOTICE
Chapter 8 Replacement procedures
LOGIQ E9 S
Corresponding Graphic
BEP Cable to Backplane J3 of GFI Configuration
(views inside the BEP)
BEP Cable to Backplane, J2 (left) of
MRX Configuration
M