GE LOGIQ E9 Service Manual page 335

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
7-5-13
Touch Panel
7-5-13-1
Calibration
NOTE:
Always try TP calibration verification before attempting to calibrate.
Verification" on page
calibrate and the normal calibration procedure does not work, it may be necessary to do a
complete Pre-calibration. See:
Upper OP FRUs" on page
To do a Touch Screen Calibration, follow the directions on the Touch Screen. As each of the cross-hairs
appear, touch them with your finger or a pencil eraser.
NOTE:
You MUST hold your finger on the cross-hair until it moves to the next location. If you just tap
the calibration cross-hair, there is a good chance your calibration will be corrupt.
7-5-13-2
Calibration Verification
The Start Touch Screen Verification brings up a grid. Each time you touch the screen a small red dots
appears where you touch, this will indicate that the calibration is correct and you do not need to
re-calibrate.
If the red dots do not show up at the touched places, you need to run the Touch Panel calibration. To
terminate the verification, move the trackball pointer into the grid and press the set key.
. 2
EV
7-87. If the Touch Panel is so far out of calibration, hangs or does not
Section 6-5 "Touch Panel Pre-Calibration when Replacing
6-12.
Chapter 7 Diagnostics/Troubleshooting