4-3-5-3
Preparations
Use a phantom (optional) when doing these tests.
1.) Connect one of the probes, to the scanner's active probe connector.
-
-
2.) Turn ON the scanner.
The B-Mode window is displayed (default mode).
3.) If needed, adjust the Display's Brightness and Contrast setting (see:
adjustments" on page
ALWAYS USE THE MINIMUM POWER REQUIRED TO OBTAIN ACCEPTABLE IMAGES
WARNING
WARNING
IN ACCORDANCE WITH APPLICABLE GUIDELINES AND POLICIES.
1.) Press B-MODE on the Operator Panel to access B-Mode.
2.) These Image Controls are used to optimize the B-Mode picture:
-
-
-
-
-
-
-
4-3-5-4
Checks
Check Width, Focus, Framerate, Frequency
The results of these adjustments must be verified on the B-Mode sector on the screen.
Check Up/Down, Left/Right, B Color Maps and Cineloop
The results of these adjustments must be verified on the B-Mode sector on the screen.
Check Gain, TGC and Depth
Check B-Mode Soft Menu Controls
Check Compress, Contour, Reject and Tilt
Check Power and Dynamic Range
4 - 34
. 2
EV
see:
Section 3-6-6 "Connecting Probes" on page 3-20
For available probes, see:
6-2).
Use Gain and TGC controls to optimize the overall image together with the Power control.
Use Depth to adjust the range to be imaged.
Use Focus to center the focal point(s) around the region of interest.
Use Frequency (move to higher frequencies) or Frame rate (move to lower frame rate) to
increase resolution in image.
Use Frequency (move to lower frequency) to increase penetration.
Use the control to optimize imaging in the blood flow regions and make a cleaner, less noisy
image.
Use Reject controls to reduce noise in the image.
Section 4-3 - Functional checks
Section 9-18 "Probes" on page
LOGIQ E9 S
ERVICE
for info about connecting the probes.
9-104.
Section 6-3 "LCD Monitor
M
ANUAL