GE LOGIQ E9 Service Manual page 180

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
4-3-7
System CFM and PWD Checks
4-3-7-1
Introduction
For information on the system's different modes as well as expected results, the Basic User Manual or
User Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
For complete information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter
5.
4-3-7-2
Preparations
Use a phantom (optional) when doing these tests.
1.) Connect one of the probes, to the scanner's active probe connector.
-
-
2.) Turn ON the scanner.
4-3-7-3
Color Mode Checks
Color Flow screens are B or M-Mode screens with colors representing blood or tissue movement. Color
Flow may be selected both from B-Mode, or from M mode, or a combination of these.
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
4-3-7-4
Color B-Mode
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
4-3-7-5
Color M-Mode
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
Check Horizontal Sweep, PRF, Baseline, and Invert
Check Variance, Color Maps and Cineloop
Check Sample Volume
Check Frequency and Acoustic Power
Adjust ROI
4 - 36
. 2
EV
see:
Section 3-6-6 "Connecting Probes" on page 3-20
For available probes, see:
Section 4-3 - Functional checks
Section 9-18 "Probes" on page
LOGIQ E9 S
ERVICE
for info about connecting the probes.
9-104.
M
ANUAL