GE
D
5535208-100, R
IRECTION
4-3-7
System CFM and PWD Checks
4-3-7-1
Introduction
For information on the system's different modes as well as expected results, the Basic User Manual or
User Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
For complete information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter
5.
4-3-7-2
Preparations
Use a phantom (optional) when doing these tests.
1.) Connect one of the probes, to the scanner's active probe connector.
-
-
2.) Turn ON the scanner.
4-3-7-3
Color Mode Checks
Color Flow screens are B or M-Mode screens with colors representing blood or tissue movement. Color
Flow may be selected both from B-Mode, or from M mode, or a combination of these.
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
4-3-7-4
Color B-Mode
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
4-3-7-5
Color M-Mode
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
•
Check Horizontal Sweep, PRF, Baseline, and Invert
•
Check Variance, Color Maps and Cineloop
•
Check Sample Volume
•
Check Frequency and Acoustic Power
•
Adjust ROI
4 - 36
. 2
EV
see:
Section 3-6-6 "Connecting Probes" on page 3-20
For available probes, see:
Section 4-3 - Functional checks
Section 9-18 "Probes" on page
LOGIQ E9 S
ERVICE
for info about connecting the probes.
9-104.
M
ANUAL