D
5535208-100, R
IRECTION
7-3-3
Collecting a Screen Capture with Logs
If the system malfunctions, press the Alt+D keys simultaneously. This Alt+D function is available at all
times, and collects a screen capture of the image monitor, user-defined presets, and the following logs:
Keyboard Shadow Log
Error Logs
Crash Log
Vital Product Data
DICOM Logs
Windows Event Logs
Diagnostic Logs
Service Logs
For a detailed list of Service Logs captured, see:
7-5.
When Alt+D is pressed, a menu box opens. Enter the following information:
System ID serial number.
Software version.
System Date and time of occurrence.
Sequence of events leading to issue.
Is the issue repeatable?
Imaging mode, probe, preset/application.
Media brand, speed, capacity, type.
Select the Destination (storage media or Service directory for remote viewing through InSite ExC)
and click the Store button.