Section 2-2

Console Requirements

2-2-1

Unit Environmental Requirements

Requirement
Operational
Storage
Transport
CAUTION: If the system has been in storage or has been transported, please see the acclimation
requirements before powering
2-2-2
Cooling Requirements
The cooling requirement for the Vivid i n ultrasound unit environment is 3500 BTU/hr. This figure does
not include the cooling required for lights, people, or other equipment in the room.
Note:
Each person in the room places an additional 300 BTU/hr demand on the environmental cooling.
2-2-3

Lighting Requirements

For system installation, updates and repairs, bright lighting is required. However, operator and patient
comfort may be optimized if the room lighting is subdued and indirect when a scan is being performed.
Therefore, a combination lighting system (dim/bright) is recommended. Keep in mind that lighting
controls and dimmers can be a source of EMI which could degrade image quality. These controls should
be selected to minimize possible interference.
2-2-4

Time and Manpower Requirements

Site preparation takes time. Begin pre-installation checks as soon as possible to allow sufficient time to
make any required changes. If possible, begin these checks as many as six weeks before system
delivery.
CAUTION: Only one person is required to unpack the Vivid i n ultrasound unit; at least two people
must be available to roll the system down the wheeling ramp. Attempts to move the system
considerable distances (or on an incline) by one person alone, could result in personal injury,
and/or damage to the system.
2-2
5
EVISION
Temperature
o
10 — 40
C
o
- 10 — 60
C
o
- 10 — 60
C
Table 2-2
Section 2-2 - Console Requirements
Relative Humidity
(non-condensing)
o
30 — 85%
(50 — 104
F)
o
30 — 95%
(14 — 140
F)
o
30 — 95%
(14 — 140
F)
Environmental Requirements
and/or using the system. See "Installation Warnings" on page 3-2.
ON
V
S
IVID I N
ERVICE
Air Pressure
700 — 1060 hPa
700 — 1060 hPa
700— 1060 hPa
M
ANUAL