Table 4-2
Step
5.
Select System to access the System Setup window.
6.
Proceed to select the required settings.
4-2-5

Using Removable Media

4-2-5-1
Introduction
Vivid i n is equipped with a Magneto Optical Disk Drive (MOD Drive) and a DVD/CD-RW Drive.
The MOD Drive can be connected to either of the USB ports (upper or lower) on the rear connectors
panel. This connection must be via a USB-to-SCSI adaptor; one side of the adaptor being connected to
the Vivid i n USB port, and the other side to the rear of the MOD drive.
When connecting the MOD drive directly to the system, it is necessary to use an additional power
source. As a safety precaution, this must be done via an isolation transformer.
The MOD Drive may be connected once the system is powered ON, or after shutdown. All software
drivers for the MOD Drive are pre-configured and installed designated to be used with the specific MOD
drive supplied by GE Healthcare.
WARNING Do not attempt to use a different type of mass storage device other than the
MOD Drive supplied with the system. The ultrasound system is an extremely sensitive and
complex medical system. Any unauthorized peripherals may cause system failure or damage!
5
EVISION
Logging on to the System as ADM (cont'd) 2 of 2
Task
Chapter 4 - General Procedures and Functional Checks
Expected Result(s)
The System Settings window is displayed
.