GE H
EALTHCARE
D
FL091021, R
IRECTION
3-8-3
Connectivity Configuration
NOTE:
If connected to a stand-alone network (Peer-to-Peer network with a Vivid i n scanner, an
EchoPAC PC work station and eventually a network printer), you should use default delivery
settings.
3-8-3-1
Introduction
To be able to use the network functions when connected to a hospital network, the scanner must have
a proper network address.
•
Before you can set up the scanner, you need to collect some information. Refer to the
for DICOM Network Information in Figure 2-2 on page 2-10 .
information is the network administrator.
•
Follow the steps below to prepare the scanner for use on the network.
3-8-3-2
Select TCP/IP Set-up Screen
1.) Press CONFIG (F2) and log on as Adm, as described in "Log On to the System as ADM" on page 4-
10.
2.) If not already selected, select CONNECTIVITY from the bottom of the screen.
3.) Select the TCP/IP TAB (it is named Tcpip).
.
Computer Name:
For Vivid i n, this name is on the form:
Vivid_I-00nnnn, where "00nnnn" is a number
(nnnn is the scanner's serial number).
IP Settings:
IP Settings area for the Vivid i n scanner.
Remote Archive Setup area:
IP settings area for a Remote Archive.
Example: EchoServer.
Default Setup:
Remote Archive IP-Addr: 10.0.0.4
Remote Archive Name: EchoPAC7-000001
Figure 3-56 TCP/IP Set-up Screen for Vivid i n, Overview (Example)
3-70
5
EVISION
Section 3-8 - Connectivity Setup
V
S
IVID I N
ERVICE
Typically, the source for this
M
ANUAL
"Worksheet