Performing the Battery Calibration Procedure (cont'd)
5.) Disconnect the AC power cable from the Vivid i n, then click Next.
The second phase of the process (Full Discharge) commences and the Battery Calibration Wizard
(Page 3) is displayed as shown in Figure 6-10 on page 6-11 .
6.) Leave the system overnight to complete the Full Discharge process.
The Vivid i n system automatically shuts down at the end of this process; the battery calibration
procedure is now complete.
7.) The following morning, re-connect the AC power cable and turn ON power to the system.
Note:
The system may now be used normally, without any dependence on battery capacity
status.
5
EVISION
Figure 6-10 Battery Calibration Wizard - Page 3
Chapter 6 - Service Adjustments
V
S
M
IVID I N
ERVICE
ANUAL
6-11