8-2-5
LCD Display Cover Hinges Replacement Procedure
8-2-5-1
Tools
Phillips screwdriver
8-2-5-2
Time Required
30 minutes
8-2-5-3
Preparations
Shut down the Vivid i n ultrasound unit, as described in
8-2-5-4
LCD Display Cover Removal Procedure
1.) Remove the Control Panel and Keyboard, as described in the
Removal Procedure" on page
2.) Disconnect the LCD Flex ribbon cable from the BEP, as shown in
Use a flat screwdriver to carefully release the cable from the connector.
3.) Loosen and remove the 2 screws from each of the LCD hinges, as shown in
8-16