IRECTION
4-3-3-4

Plug and Scan Test

Table 4-17
Press the On/Off button on the control console for no more
1
than 3 seconds.
2
Turn OFF the system's circuit breaker.
Turn ON the system's circuit breaker.
3
Press the On/Off Button.
Turn OFF the system's circuit breaker. Turn ON the
4
system's circuit breaker after entering to Standby mode.
Turn OFF the system's circuit breaker.
5
Press the On/Off button.
5
EVISION

Plug and Scan Test - Recommended

Task
Chapter 4 - General Procedures and Functional Checks
V
IVID I N
Expected Result(s)
The system enters Standby mode.
The system remains in Standby mode, indicated by the
blinking On/Off button LED.
The system returns quickly to the Power-ON state -less
than 30 seconds.
The system enters Standby mode and stays in this mode.
The On/Off button LED is blinking.
The system enters the Power-OFF state. The On/Off
button LED is OFF.
S