Removing The Probe And Mux Board From The Bottom Assembly; P&M (Probe And Mux) Board Installation Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
8-3-1-4
P&M (Probe and MUX) Board Removal Procedure (cont'd)
4.) Lift the P&M Board upwards away from the Bottom Assembly and remove it completely, as shown
in
Figure 8-24
Figure 8-24 Removing the Probe and MUX Board from the Bottom Assembly
8-3-1-5
P&M (Probe and MUX) Board Installation Procedure
CAUTION: When installing boards, ESD may cause damage to a board. Always have the ESD hand
strip connected to the machine chassis and to your hand.
1) Return a new P&M Board to the system, pushing it downwards into position on the Bottom
Assembly. Make sure it is properly seated before returning and fastening the two screws (previously
removed). Refer to
2.) Return and fasten the four screws to secure the P&M Board to the RFI Board, as shown in
Figure 8-22 on page 8-25 .
3.) Return the Control Panel and Keyboard, as described in the
Installation Procedure" on page 8-8
4.) Proceed to perform the following functionality tests:
-
-
-
8-26
5
EVISION
below.
Figure 8-24
"Chassis Current Leakage Test" on page 10-25
"Calibration Tests" on page 7-15
"Full System Test" on page 7-9
Section 8-3 - Internal Component Replacement Procedures