GE Vivid i n Service Manual page 83

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
3-6-1-4
DVD/CD-RW Drive (Older Type) (cont'd)
Note:
When plugging the DVD's AC adaptor directly into the mains supply, there is no isolation
from leakage currents since the USB connector is not isolated. It is therefore necessary to
use an isolation transformer to isolate the mains power supplied to the DVD's AC adaptor.
If an isolation transformer is not available, the DVD must be disconnected from the
Vivid i n system while scanning a patient.
The DVD/CD-RW Drive can be connected once the system is powered ON, or after shutdown. All
software drivers for the DVD/CD-RW Drive are pre-configured and installed designated to be used
with the specific DVD/CD-RW supplied with the system.
WARNING Do not attempt to use a different type of mass storage device other than the
DVD/CD-RW Drive supplied with the system. The ultrasound system is an extremely sensitive and
complex medical system. Any unauthorized peripherals may cause system failure or damage!
5
EVISION
Chapter 3 - System Setup
V
S
M
IVID I N
ERVICE
ANUAL
3-25

Advertisement

Table of Contents
loading

This manual is also suitable for:

H45561acH45551deFl000110Fl000090

Table of Contents