IRECTION
2-3-4

Networking Pre-Installation Requirements

2-3-4-1
Stand-alone Unit (without Network Connection)
None.
2-3-4-2
Unit Connected to Hospital's Network
Supported networks:
Wired LAN
Wireless LAN
2-3-4-3

Purpose of the DICOM Network Function

DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote printers. As an added benefit, transferring
images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while
scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at
a lower cost.
2-3-4-4

DICOM Option Pre-Installation Requirements

To configure the Vivid i nultrasound unit to work with other network connections, the network
administrator must provide the required information, which should include the following:
Vivid i n details:
Routing Information:
DICOM Application Information:Details of the DICOM devices in use at the site, including the
5
EVISION
Chapter 2 - Site Preparations
DICOM network details for the Vivid i nunit, including the host
name, local port, IP address, AE title and net mask.
IP addresses for the default gateway and other routers in use at
the site.
DICOM host name, AE title and IP addresses.
V
S
M
IVID I N
ERVICE
ANUAL
2-9