Peripheral/Option Checks; Input Power Checks; Mains Cable Inspection; Ge Approved Peripheral/Hardware Option Functional Checks - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
10-5-2-2

Peripheral/Option Checks

If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to
the Vivid i n for a list of approved peripherals/options.
Table 10-6
Step
1
2
3
4
5
6
7
8
10-5-3

Input Power Checks

10-5-3-1

Mains Cable Inspection

Table 10-7
Step
1
2
3
4
10-10
5
EVISION

GE Approved Peripheral/Hardware Option Functional Checks

Item
VCR
Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
Verify that InSite is functioning properly. Ensure two-way remote communications.
InSite/iLinq
(Warranty & Contract Customers only)
ECG
Verify basic operation with customer.
DVD-CDRW
Verify that the DVD-CDRW is functioning properly.
MOD
Verify that the MOD is functioning properly.
Mains Cable Inspection
Item
Unplug Cord
Disconnect the mains cable from the wall and system.
Mains Cable
Inspect the mains cable and its connectors for any damage.
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and
Terminals
that no strands may cause a short circuit.
Inlet Connector
Verify that the Inlet connector retainer is functional.
Section 10-5 - System Maintenance