External Component Replacement Procedures; Battery Replacement Procedure; Battery Removal Procedure; Closing Vivid I N Lcd Cover - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
Section 8-2

External Component Replacement Procedures

8-2-1

Battery Replacement Procedure

8-2-1-1
Tools
None required.
8-2-1-2
Time Required
1 minute
8-2-1-3
Preparations
Shut down the Vivid i n ultrasound unit, as described in
8-2-1-4

Battery Removal Procedure

1) Disconnect the AC adapter and all cables from the Vivid i n.
2) Close the Vivid i n LCD cover, making sure both front latches are securely fastened.
Latch
3) Carefully turn the Vivid i n over, placing it face down.
4.) While holding the Vivid i n steady, pull the locking latch away from the battery and slide the battery
sideways out of the battery compartment to remove it completely from the system.
WHEN REMOVING A DEFECTIVE BATTERY, ENSURE THAT IT IS DISPOSED OF IN
NOTE:
ACCORDANCE WITH LOCAL REGULATIONS. ALTERNATIVELY, FORWARD IT TO
GE Healthcare
8-2
5
EVISION
Figure 8-1 Closing Vivid i n LCD Cover
Figure 8-2 Sliding the Battery out of the Battery Compartment
FOR PROPER DISPOSAL.
Section 8-2 - External Component Replacement Procedures
V
IVID I N
Chapter 3
-
System
Setup.
Latch
S
M
ERVICE
ANUAL