3-6-3

Connecting Probes

The Vivid i n ultrasound unit operates with various types of probes that are used for scanning patients,
including flat phased, convex and linear electronic array probes. Once connected, the probes can be
selected for different applications.
Probe connectors on the Vivid i n unit are as follows:
Two active probe connectors (one for a pencil probe) on the right side of the unit
Probes can be connected or changed any time, as described below, regardless of whether the system
is powered ON or OFF.
CAUTION: Handle the probe gently while connecting and disconnecting.
Do NOT touch the patient and any of the connectors on the ultrasound unit simultaneously, including
ultrasound probe connectors.
1) Place the probe's carrying case on a stable surface and open the case.
2) Inspect the probe socket to verify that it is free of debris.
3) Carefully remove the probe and unwrap the probe cord.
DO NOT allow the probe head to hang freely. Impact to the probe head could result in irreparable
NOTICE
damage.
4) Press the probe connector locking lever downwards (refer to
3-40
5
EVISION
Figure 3-27 Probe Connection Locking Lever
Section 3-6 - Completing the Hardware Installation
V
S
IVID I N
ERVICE
Figure
3-27).
M
ANUAL