Microphone Assembly Replacement Procedure
8-3-9-1
Tools
None required.
8-3-9-2
Time Required
10 minutes
8-3-9-3
Preparation
Shut down the Vivid i n ultrasound unit, as described in
8-3-9-4
Microphone Assembly Removal Procedure
1) Remove the LCD Display Frame as described in the
page
2) Carefully unclip and release the microphone from the upper right side of the LCD display, as shown
in
Figure 8-53
3) Release the microphone cable from its seating in the cable groove that runs along the side and
bottom of the LCD Display.
Note:
An adhesive rubber channel is fitted inside the groove to hold the cable in place.
4.) Disconnect the miniature plug from the cable connector below the LCD Display as shown in
8-54.
5.) Remove the Microphone Assembly from the system.
8-54
5
EVISION
8-22.
below.
Figure 8-53 Releasing the Microphone
Figure 8-54 Disconnecting the Miniature Plug
Section 8-3 - Internal Component Replacement Procedures
V
IVID I N
Chapter
3-
System
Setup.
"LCD Display Frame Removal Procedure" on
S
M
ERVICE
ANUAL
Figure