GE Vivid i n Service Manual page 409

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
8-4-7
Full Re-ghost and Software Installation (cont'd)
After re-ghosting (requires approximately another 10 minutes), the software installation will be
complete. The system will re-start and prompt for the software License Password.
9) Type in the password provided for the specific software, and approve by clicking OK.
A.
B.
C.
10) Restart the system and commence scanning.
11.)Proceed to perform the following functionality tests:
-
-
-
5
EVISION
Install Software
Service Tools
Exit
"Software Configuration Checks" on page 4-28
"Calibration Tests" on page 7-15
"Full System Test" on page 7-9
Chapter 8 - Replacement Procedures
V
S
M
IVID I N
ERVICE
ANUAL