GE H
EALTHCARE
D
FL091021, R
IRECTION
7-3-1-1
Full System Test (FE and BEP) (cont'd)
4.) Adjacent to On Fail, select either Stop or Continue, as required. Stopping on failure enables you
to review a failed test, the moment it fails. Alternatively, you may prefer testing to continue on failure
and later you may review the report.
5.) Trackball to the Start button and press Set.
The Automatic Test Sequence commences. While the sequence of automatic testing or the group
of tests proceeds, the name of the test currently in progress (Full System Test FE and BEP) is
displayed in the Execute field and the current Loop # and specific test are shown below. The
messages Init Done and Please Wait are displayed beside the progress bars in the Status area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full System Test (FE and BEP) sequence is complete:
-
-
7-12
5
EVISION
Green:
All recommended tests for these components have been completed and no
problems were identified.
Red:
Problems were identified in these components during the performance of the test.
Yellow:
The tests executed so far on these components have passed, but not all
recommended tests have been performed.
the Diagnostic Test window displays the Finished message
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in
Section 7-3 - Performing Full System Diagnostics